Products

AMS2750 Approved Vacuum Furnace Thermocouple with Form 'A' Flip Top Die Cast Alloy Head

AMS2750 Approved Thermocouples for Vacuum Furnaces
with Form 'A' Flip Top Die Cast Alloy Head

Ceramic Sheathed Vacuum Thermocouples
Type 40/41 assemblies incorporate an integral Spectite® seal and are ideal for vacuum furnace applications. The Spectite® vacuum seal is rated to 5 x 10-6 torr with low leak rates (better than or equal to 1 x 10-6 scc/sec. under 1 Atm @ 20ºC). These sensors can be supplied to all recognised standards and tolerances when combined with our UKAS calibration services. They are available in simplex and duplex junction styles with an optional test sensor hole for calibration / survey work. A wide range of support tube fittings for attachment into the process is available.
  • Vacuum seal rated 5 x 10-6 torr
  • Low leak rates
  • Integral Spectite® seal on internal wire maintains vacuum even if sheath is damaged
  • Wide operating temperature range up to 2910°F
  • Optional test sensor hole for calibration or system accuracy tests (SAT) without removal from process
  • Wide range of attachments for process installation
  • Simplex and Duplex versions available
  • UKAS and In-House calibration options available to all major international and industry specific standards
  • Flexible tagging options available for clear traceability
Calibration Accuracies:

Code Standard Accuracy Supplied Certification
IEC1 IEC 60584.1 Class 1 Types K/N: ±1.5ºC (-40 to +375ºC) or 0.4% (375 to 1000ºC)
Types R/S: ±1.0ºC (0 to 1100ºC) or 1ºC + 0.3% of temperature above 1100ºC (1100 to 1600ºC)
TC offer both UKAS and In-House (traceable to national standards) calibrations to all major international standards up to 1590°C. Our laboratory is fully accredited to ISO17025 and our reports are designed with the requested standard in mind. Our sales engineers are on hand to discuss any requirements so please do not hesitate to contact us for more details.
AMS1 AMS2750 Types K/N: ±1.1ºC or 0.4%
Types R/S: ±1.0ºC or 0.25%
(whichever is greater)
Type B: ±1.0ºC or 0.25%
(whichever is greater)
BAC2 BAC5621 ±1.1ºC <538ºC or
0.4% of reading >538ºC



Vacuum Thermocouple Enquiry Form

1. Thermocouple Type
2. Sheath Material
3. Types of Sensing Junction

Illustration Select
INSULATED Junction
EXPOSED Junction
Duplex Element Required? (please tick box)
4. Sheath Diameter
5. Sheath Length (L)
 inches
6. Support Tube Length (L1)
 inches
7. Optional Test Sensor Hole

Illustration Select
Hole supplied for test sensors up to 0.062" diameter
(check box if required)
8. Optional Sheath Fitting

Illustration Select
Vacuum Feedthrough with Viton Sealant
1/4" NPT
1/2" NPT
3/4" NPT
Welded Fixed Position Screwed Bush
3/4" NPT
1" NPT
Welded Vacuum Flange
Welded KF vacuum flange
Welded CF (ConFlat) vacuum flange
9. Tagging Options

Illustration Select
Plastic ID Label
For use with all end seals up to 158ºF
Laser Etched Metal Tag
For use with all end seals above 480°F
Laser Etching on Sensor
Laser etch of serial number. This can be instead of or in addition to either of the options above
10. Calibration Accuracies

Calibration Type Select
IEC1 - IEC 60584.1
Class 1
AMS1 - AMS2750
BAC2 - BAC5621
Your chosen calibration is:

IEC1- IEC 60584.1 Class 1
Accuracy:
Types K/N: ±1.5ºC (-40 to +375ºC) or 0.4% (375 to 1000ºC)
Types R/S: ±1.0ºC (0 to 1100ºC) or 1ºC + 0.3% of temperature above 1100ºC (1100 to 1600ºC)
Certification:
TC offer both UKAS and In-House (traceable to national standards) calibrations to all major international standards up to 1590°C. Our laboratory is fully accredited to ISO17025 and our reports are designed with the requested standard in mind. Our sales engineers are on hand to discuss any requirements so please do not hesitate to contact us for more details.
Your chosen calibration is:

AMS1 - AMS2750
Accuracy:
Types K/N: ±1.1ºC or 0.4%
Types R/S: ±1.0ºC or 0.25%
(whichever is greater)
Type B: ±1.0ºC or 0.50%
(whichever is greater)
Certification:
TC offer both UKAS and In-House (traceable to national standards) calibrations to all major international standards up to 1590°C. Our laboratory is fully accredited to ISO17025 and our reports are designed with the requested standard in mind. Our sales engineers are on hand to discuss any requirements so please do not hesitate to contact us for more details.
Your chosen calibration is:

AMS2 - AMS2750
Accuracy:
±2.2ºC or 0.75% of reading
(whichever is greater)
Certification:
TC offer both UKAS and In-House (traceable to national standards) calibrations to all major international standards up to 1590°C. Our laboratory is fully accredited to ISO17025 and our reports are designed with the requested standard in mind. Our sales engineers are on hand to discuss any requirements so please do not hesitate to contact us for more details.
Your chosen calibration is:

BAC2 - BAC5621
Accuracy:
±1.1ºC <538ºC or 0.4% of reading >538ºC
Certification:
TC offer both UKAS and In-House (traceable to national standards) calibrations to all major international standards up to 1590°C. Our laboratory is fully accredited to ISO17025 and our reports are designed with the requested standard in mind. Our sales engineers are on hand to discuss any requirements so please do not hesitate to contact us for more details.
11. Calibration Standard
12. Report Type

Details Select
Standard 3 Page Report
An in-house 3-page report for the start/end of batch from 32-2190ºF with all the information NADCAP auditors require (please refer to the Calibration Details page for a detailed explanation)
Custom Report
As the Standard Report but with your choice of range and interval
Your chosen Report Type is:

An in-house 3-page report for the start/end of batch from 32-2190ºF with all the information NADCAP auditors require (please refer to the Calibration Details page for a detailed explanation).
Your chosen Report Type is:

A Custom Report
Please enter the calibration range:
Please enter the calibration interval:
13. Quantity Required

Contact Details

Company Name


Your Name


Title/Dept.


Email


Telephone


Additional Information

If your enquiry is for more than one style of sensor, please email us with your requirements or use the box below



Send Enquiry  


BACK TO TOP
Products

Contact Us

About TC

FREE Guide

Privacy Policy

Terms and Conditions

Sitemap
FREE Wallchart Guide

Join over 500,000 engineers and buyers who have already received our FREE 74 page A3 Guide to Thermocouple and Resistance Thermometry

Company Name
  Address 1
Your Name
  Address 2
Title / Dept.
  State
Email
  ZIP Code
Telephone
  Country
Company Name
Your Name
Title / Dept.
Address 1
Address 2
State
ZIP Code
Country
Email
Telephone

We will only ever use the information you provide to occasionally contact you regarding our products and we will never disclose your details to a third party. Our privacy policy can be viewed here



I agree to the terms and conditions listed in the Privacy Policy

View our ISO 90001/2008 Certficate

© 2006- TC Measurement and Control Inc., PO Box 685, Hillside, IL 60162, United States
Tel.: 877 249 1777, Fax: 708 449 0777, Email: info@tc-inc.com



1 Thermocouple Type Temperature Range (continuous) Temperature Range
(short term)
K Nickel Chromium vs
Nickel Aluminium
32°F to +2010°F -350°F to +2460°F
N Nicrosil vs Nisil 32°F to +2190°F -450°F to +2370°F
R Platinum - 13% Rhodium vs
Platinum
0°C to +1600°C -50°C to +1750°C
S Platinum - 10% Rhodium vs
Platinum
0°C to +1550°C -50°C to +1700°C
B Platinum - 30% Rhodium vs
Platinum 6% Rhodium
+100°C to +1600°C +100°C to +1820°C
Temperature ranges refer to the thermocouple type.
The temperature range of the sensor may deviate from the above,
please refer to the product description for more details
2 Sheath Material Operational Properties Max.
Temperature
IAP Impervious Aluminous Porcelain Ideally suited for use with base metal thermocouple conductor combinations. Has a very low temperature coefficient of expansion thus giving excellent resistance to thermal shock. Offers high strength and high resistance to flux and slag attack. Suited to kiln applications where low contamination requirements preclude the use of a metal sheath. NB. Requires support at high temperature if horizontal. 1400°C
IRA Impervious Recrystallised Alumina Ideally suited for use with precious metal thermocouples at high temperatures. Provides a fair resistance to thermal shock. High degree of inertness to chemicals. Ideal for reducing carbonaceous atmospheres and offers a high resistance to alkaline and other fluxes. 1600°C

INSULATED
Hot junction insulated from sheath. Gives floating output with typical insulation resistance in excess of 100 megohms

EXPOSED
Fastest response mainly for the measurement of air temperature in ducts. Restricted to a maximum operating temperature of 300ºC

3 Ceramic Sheath
Diameter 'd1'
Support Tube
Diameter 'd2'
Style
  3.0mm 6.0mm Exposed Junction (simplex)
  3.5mm 6.0mm Exposed Junction (duplex)
  4.0mm 6.0mm Exposed or Insulated Junction
  6.0mm 10.0mm Exposed or Insulated Junction
  8.0mm 12.0mm Exposed Junction
  10.0mm 15.9mm Exposed Junction
  12.0mm 15.9mm Exposed Junctions